Tuttnauer Autoclave Fault Codes - Complete Guide

⚡ Quick Answer

Tuttnauer autoclave fault codes for 2340, 3870, Elara, and Valueklave series: error codes, causes, and troubleshooting steps.

Tuttnauer Autoclave Fault Codes - Quick Reference

Tuttnauer autoclaves (2340, 3870, Elara, Valueklave, and EZ10 series) display fault codes on the digital or LED display panel. Common in dental, veterinary, and laboratory settings.

CodeModelMeaningQuick Fix
E1AllTemperature sensor faultCheck sensor wiring and resistance
E2AllPressure sensor faultCheck transducer signal
E32340/3870Temperature not reached in timeCheck steam supply and door seal
E4AllOverpressure safety faultCheck safety valve
E5AllDoor not closed or lockedCheck door and locking mechanism
E6ElectronicMemory/EEPROM faultContact service
E7ElaraWater level faultCheck water reservoir
E8ElaraConductivity faultCheck water quality (use distilled)
E9EZ10Printer faultCheck printer paper/connection
LODisplayLow water levelRefill with distilled water

Most Common Faults

E1 - Temperature Sensor Fault

Tuttnauer uses PT100 temperature sensors in the chamber. An E1 fault means the sensor is reading out of range - either a broken wire (reads too high) or a short (reads too low). Disconnect the sensor and test resistance: 100 ohms at 0°C, approximately 138 ohms at 100°C. Replace the sensor if values are incorrect.

E3 - Temperature Not Reached

If the autoclave doesn’t reach the target temperature (121°C or 134°C) within the preset time, E3 aborts the cycle. This is commonly caused by: door not sealing fully (check gasket), low steam generation from a scaled heating element, or insufficient water. Descale the unit if limescale buildup is visible.

E5 - Door Fault

Tuttnauer door faults indicate the door locking mechanism didn’t complete. On manual-lock models, ensure the door handle is fully rotated to the locked position. On automatic-lock models, check the door solenoid and locking cam. A worn door gasket that lets steam escape also prevents the door from locking under pressure.

E7 / LO - Water Level Fault

Elara and EZ10 models have internal water reservoirs with level sensors. Always use distilled or RO water - tap water causes mineral scaling that damages the heating element and clogs the strainer. Clean the water reservoir and level sensor if false water level alarms occur with a full reservoir.

E8 - Conductivity Fault (Elara)

Elara models test water conductivity to prevent mineral buildup damage. If tap water or inadequately treated water is used, E8 triggers and the cycle is aborted. Use only distilled or deionized water with conductivity below 15 µS/cm.

Parts Often Needed

PartNotes
PT100 temperature sensorAmazon | Replace on E1
Door gasket (silicone)Amazon | Replace on E5 / poor sealing
Heating elementAmazon | Replace on slow heat-up
Safety valveAmazon | Replace on overpressure fault
Level sensorAmazon | Replace on false LO alarm

When to Call a Pro

Tuttnauer autoclave spore testing, cycle validation, and annual PM require trained biomedical technicians. In dental and medical practices, sterilizer validation records are required by regulatory bodies.