Tuttnauer Autoclave Fault Codes - Complete Guide

⚡ Quick Answer

Tuttnauer autoclave fault codes for 2340, 3870, Elara, and Valueklave series: error codes, causes, and troubleshooting steps.

⚡ AT A GLANCEScroll down for full repair guide
What it meansTuttnauer fault codes indicate temperature sensor failures, pressure faults, door locking issues, water level problems, or conductivity errors on Elara models.
Most likely causeFailed PT100 temperature sensor, scaled heating element preventing target temperature, worn door gasket, low water level, tap water conductivity exceeding limits.
First checkTest PT100 sensor resistance (100 ohms at 0°C), inspect door gasket seal, verify distilled water in reservoir, check safety valve operation, measure water conductivity.
⏱Repair time15 minutes for water refill and gasket inspection, 1-2 hours for sensor or heating element replacement depending on model accessibility.
🔧Part neededPT100 temperature sensors, door gaskets, heating elements, safety valves, level sensors

Parts to price first

Check the part before you book service.

PT100 Temperature Sensor Probe

E1 temperature sensor fault requires PT100 replacement; test resistance (100 ohms at 0°C, 138 ohms at 100°C) before ordering to confirm sensor failure.

Autoclave Door Gasket Silicone Seal

E3 and E5 door faults trace to worn gaskets preventing proper seal; replace if gasket shows cracks, tears, or prevents door from locking under pressure.

Distilled Water for Autoclaves

E8 conductivity faults and E7 water level issues require distilled or RO water below 15 µS/cm; tap water causes mineral scaling and damages heating elements.

Autoclave Heating Element Replacement

E3 slow heat-up faults trace to scaled or failed heating elements; descale first, replace element if cycle time remains excessive after cleaning.

$85-$185Check price

Autoclave Safety Valve

E4 overpressure faults indicate safety valve failure or blockage; replace valve if manual test shows it does not open at rated pressure.

Autoclave Descaling Solution

Limescale buildup causes E3 temperature faults and slow cycles; descale quarterly in hard water areas to maintain heating element efficiency.

Equipment service manual download (eManualOnline)

The full OEM-style workshop manual has the wiring diagram, test values, and the fault table this guide summarizes. Instant PDF download.

Repair intent

Need this fixed instead of diagnosed?

Tuttnauer autoclave fault codes for 2340, 3870, Elara, and Valueklave series: error codes, causes, and troubleshooting steps.

Tuttnauer Autoclave Fault Codes - Quick Reference

Tuttnauer autoclaves (2340, 3870, Elara, Valueklave, and EZ10 series) display fault codes on the digital or LED display panel. Common in dental, veterinary, and laboratory settings.

CodeModelMeaningQuick Fix
E1AllTemperature sensor faultCheck sensor wiring and resistance
E2AllPressure sensor faultCheck transducer signal
E32340/3870Temperature not reached in timeCheck steam supply and door seal
E4AllOverpressure safety faultCheck safety valve
E5AllDoor not closed or lockedCheck door and locking mechanism
E6ElectronicMemory/EEPROM faultContact service
E7ElaraWater level faultCheck water reservoir
E8ElaraConductivity faultCheck water quality (use distilled)
E9EZ10Printer faultCheck printer paper/connection
LODisplayLow water levelRefill with distilled water

Most Common Faults

E1 - Temperature Sensor Fault

Tuttnauer uses PT100 temperature sensors in the chamber. An E1 fault means the sensor is reading out of range, either a broken wire (reads too high) or a short (reads too low). Disconnect the sensor and test resistance: 100 ohms at 0°C, approximately 138 ohms at 100°C. Replace the sensor if values are incorrect.

E3 - Temperature Not Reached

If the autoclave doesn’t reach the target temperature (121°C or 134°C) within the preset time, E3 aborts the cycle. This is commonly caused by: door not sealing fully (check gasket), low steam generation from a scaled heating element, or insufficient water. Descale the unit if limescale buildup is visible.

E5 - Door Fault

Tuttnauer door faults indicate the door locking mechanism didn’t complete. On manual-lock models, ensure the door handle is fully rotated to the locked position. On automatic-lock models, check the door solenoid and locking cam. A worn door gasket that lets steam escape also prevents the door from locking under pressure.

E7 / LO - Water Level Fault

Elara and EZ10 models have internal water reservoirs with level sensors. Always use distilled or RO water. Tap water causes mineral scaling that damages the heating element and clogs the strainer. Clean the water reservoir and level sensor if false water level alarms occur with a full reservoir.

E8 - Conductivity Fault (Elara)

Elara models test water conductivity to prevent mineral buildup damage. If tap water or inadequately treated water is used, E8 triggers and the cycle is aborted. Use only distilled or deionized water with conductivity below 15 µS/cm.

Parts Often Needed

PartNotes
PT100 temperature sensorReplace on E1
Door gasket (silicone)Replace on E5 / poor sealing
Heating elementReplace on slow heat-up
Safety valveReplace on overpressure fault
Level sensorReplace on false LO alarm

Need Replacement Parts?

When to Call a Pro

Tuttnauer autoclave spore testing, cycle validation, and annual PM require trained biomedical technicians. In dental and medical practices, sterilizer validation records are required by regulatory bodies.

Quick FAQ

What does Tuttnauer E1 error code mean and how do I fix it?

E1 indicates a PT100 temperature sensor fault from broken wiring or sensor failure. Disconnect the sensor and test resistance with a multimeter (100 ohms at 0°C). Replace the sensor if values are incorrect or if you see an open circuit.

Why does my Tuttnauer autoclave show E3 temperature not reached?

E3 occurs when the unit cannot reach 121°C or 134°C in the preset time. Check door gasket seal, descale the heating element if limescale is visible, verify water level, and ensure steam generation is adequate.

How do I fix Tuttnauer E8 conductivity fault on Elara models?

E8 triggers when water conductivity exceeds 15 µS/cm. Use only distilled or deionized water. Drain and refill the reservoir with proper water to clear the fault.

When should I call a biomedical technician for autoclave errors?

Call for spore testing, cycle validation, annual preventive maintenance, or if faults persist after basic troubleshooting. Medical and dental practices require documented validation records for regulatory compliance.